VQPet™ H Blood Glucose Monitoring System For Pet Use

Calibrated for Cats, Dogs, Rabbits and Rats

  • Easy test with one button operation to test blood glucose for pets at home
  • Only 0.5 μL blood sample needed
  • One meter can be used for testing cats, dogs, rabbits and rats
  • Glucose dehydrogenase-flavin adenine dinucleotide (GDH-FAD) enzyme used
  • Fresh whole blood (capillary and venous) can be used for testing for veterinary clinics

Advanced features for 8 electrodes strip

  • Code calibration (auto-coding)
  • Check sample sufficient or not
  • Eliminates the HCT (hematocrit) interference
  • Clucose reaction signal detection
  • Check test strip damaged or not
  • Check the humidity exposure

Specification

Test range 10-600 mg/dL (0.6-33.3 mmol/L)
Result calibration Plasma-equivalent
Test sites Dogs: inner lip and ear
Cats: callus of leg, ear and paw
Rabbits: ear
Rats: eye socket and tail
Sample type Fresh whole blood(capillary, venous)
Enzyme Glucose dehydrogenase
Sample size About 0.5 μL
Testing time About 5 seconds
Operation temperature 5°C - 45°C (41°F - 113°F)
Operation humidity 10% - 90%
Hematocrit range 0-70%
Battery One (1) CR 2032 3.0V coin cell battery
Battery life Over 1,000 tests
Memory 300 results with time and date

VQPet™ H Blood Glucose Monitoring System For Pet Use

Calibrated for Cats, Dogs, Rabbits and Rats

  • Easy test with one button operation to test blood glucose for pets at home
  • Only 0.5 μL blood sample needed
  • One meter can be used for testing cats, dogs, rabbits and rats
  • Glucose dehydrogenase-flavin adenine dinucleotide (GDH-FAD) enzyme used
  • Fresh whole blood (capillary and venous) can be used for testing for veterinary clinics

Advanced features for 8 electrodes strip

  • Code calibration (auto-coding)
  • Check sample sufficient or not
  • Eliminates the HCT (hematocrit) interference
  • Clucose reaction signal detection
  • Check test strip damaged or not
  • Checks the humidity exposure
Specification
Test range 10-600 mg/dL (0.6-33.3 mmol/L)
Result calibration Plasma-equivalent
Test sites Dogs: inner lip and ear
Cats: callus of leg, ear and paw
Rabbits: ear
Rats: eye socket and tail
Sample type Fresh whole blood (capillary,venous,arterial,neonatal)
Enzyme Glucose dehydrogenase
Sample size About 0.5 μL
Testing time About 5 seconds
Operation temperature 5°C - 45°C (41°F -113°F)
Operation humidity 10%-90%
Hematocrit range 0-70%
Battery One (1) CR 2032 3.0V coin cell batteries
Battery life Over 1,000 tests
Memory 300 results with time and date